Sm-g891a cs3 rd1 combination file download






















Filling in of Site specific Data Checking Software Version Cleaning Image Plates before first Use Technical Image Quality Check of the System Creating the Flatfield Test Image Importing the Limit Pattern Checking the Images at the Lightbox Checking the UPS Option Hand-over of the System to the Clinical Application Specialist NOTE: The Installation Checklist in section 4 of this chapter gives an overview of all steps and guides through the complete installation process.

Installation Prerequisites The purpose of this section is to show the required tasks before installation to guarantee a smooth digitizer installation. The following table lists these tasks:. Check of Installation Site Prerequisites 1 Confirm that all prerequisites described in the "Installation readiness checklist" are fulfilled.

This checklist is available at the end of Chapter 11, Installation Planning. Digitizer Unpacking 1 Compare the labels on the boxes with the customer's order list and the shipping papers. If damage is visible from the outside contact your local AGFA representative. Unpack the digitizer. The unpacking instructions are printed on the digitizer box. Figure 2 next page shows the unpacking instructions. In case it is not complete inform the national service manager. He will decide how to proceed.

Risk of injuries when lifting the digitizer. At least three people have to lift the digitizer from one position to the other. Use proper shoe and hand protection when lifting the digitizer. The structure and stability of the used table need to be suitable in relation with the size and weight of the system. The table may not be subject to excessive shock and vibrations from other sources.

Figure 8 4 Adapt dip switches to available output voltage. Refer to the UPS user's guide which is part of delivery. When connecting digitizer and control PC at two different sockets, installed in the wall: Inform the hospital technician that digitizer and control PC have to be connected to the same ground and the same phase.

Fasten the FireWire cable in the digitizer rear at the cable holder left or right. IMPORTANT: Do not place the laptop on top of the digitizer, because the accumulated heat developed by the digitizer and the laptop may affect the operation of the system. Installation The purpose of this section is to show the required installation tasks. NX workstation, in its function to have the NX image processing software running on it. If the NX workstation is operational it takes approx.

See Table 1 for meaning of the status indicator colors. Error will be solved by the registration of the digitizer serial number on the control PC as described in this section. Error if present will be solved with a check of the installed software version, see step 2. The Software Version loaded onto the digitizer and that one installed on PC do not match.

Please perform the upload via Service Client. Only displayed if software on PC and digitizer do not match. The steps outlined here are only to confirm that the NX workstation was prepared by production for the CR X system.

See also important note below. Checking Software Version NOTE The digitizer only boots-up completely if the software version on the digitizer and the software version on the control PC for download to the digitizer match.

As digitizer and control PC are not staged together, it is possible, that the digitizer or the control PC has a different digitizer software version installed. In this case error appears with each re-boot of the control PC. For installation instructions refer to section 3. No further action required.

Exception: If newer digitizer software is available on MEDNET in the freeware area perform digitizer software update or upgrade according to the installation instruction which is part of the software. Completion of Installation The purpose of this section is to show the tasks that are required to prepare the digitizer for hand-over to the clinical application specialist. The following table lists these tasks. Do not take an image plate of one tray and put it in another tray.

Use only a dry cloth or blow away possible dust with compressed air to clean the tray. When ordering, the actual revision of the spare part is delivered.

Open the shutter. Turn the cassette around, so that the black tube side is above. Let the black tray and image plate slide out carefully onto the table. Put the black tray with the image plate onto the cassette.

The tray has 2 little hooks. Put the tray onto the cassette in such a way that the hooks reach over the border of the cassette. This to avoid a bending of the tray.

Check the image plate for scratches. Clean the image plate and the tray as follows:. Moisten the cloth with cleaner and wipe the image plate surface softly and evenly phosphorous side and rear side.

Leave the image plate surface to dry for about 10 minutes to allow the solvents to evaporate. Use only a dry cloth or blow away possible dust with canned air to clean the tray thoroughly. Once the image plate surface is dry after about 10 minutes , check once again for particles of material and other impurities.. Put the tray with the image plate back into the cassette. Technical Image Quality Check of the System During the Technical Image Quality Check of the system a flatfield image is exposed and compared at the lightbox with a test pattern lowest acceptable quality.

This procedure is divided in 3 steps:. Compare the scanned flatfield with the imported limit pattern lowest acceptable quality. Creating the Flatfield Test Image 1 Select the largest available cassette. If possible, use a 35x43 cm Genrad cassette. NOTE: Select the following exposure parameters using a 1. The values are approximate and may vary among the X-ray devices. Turn the cassette by Expose again with 10 Gy.

Insert the exposed cassette in the digitizer. Any other examination type may give a result which cannot be compared with the limit pattern. Checking the Images at the Lightbox 1 At the lightbox or the viewing station visually compare the scanned flatfield with the limit pattern. The limit pattern represents the lowest acceptable image quality in slow scan and fast scan direction.

If the scanned image visually looks as good or better than the limit pattern, the image quality is okay. If there are unacceptable effects in slow scan or fast scan direction see chapter 3.

NOTE: The slow scan direction is always the long side of the image plate. These darker zones also have to be used for image quality evaluation in slow scan and fast scan direction.

See next page. Flatfield with Genrad Cassette. It shows the most important steps of the last scan cycle. Saving the HOT file documents the successful completion of the technical installation of the digitizer as well as the successful connection to the control PC.

Hand-over of the System to the Clinical Application Specialist 1 Inform the clinical application specialist about completion of technical installation. System Overview CRX in a Nutshell NX Workstation in a Nutshell Image Plates with IPs in a Nutshell Digitizer Boot-Up Functions of Calibration Board Signal Corrections before and during Scan Stall Detection Explanation of one scan cycle: 1 The user inserts the cassette with image plate IP into the digitizer 2 The chip reader reads the image plate data 3 The user enters the patient data at the the NX workstation and selects the ID button to send the data to the digitizer 4 The cassette unit opens the cassette and pulls out the tray with IP 5 The transport unit drives the image plate to the slow scan unit 6 Via calibration board the digitizer determines the position of the polygon mirror to synchronize scanning and digitizing.

The processed images are stored on the internal hard disk of the NX Workstation. However, the NX Workstation can only provide a limited amount of storage capacity. The final archiving must be realized either by printed hardcopies or a PACS.

Moreover, the NX Workstation decodes the routing data and sends the image data to the selected destinations e. Printer, PACS. For the NX Workstation a lot of licensed software options e. The tray is used as carrier to pull out the image plate and put it back into the cassette. The tray only leaves the cassette for approx. Figure 4 Workflow of an examination with the CRX system in the view of a radiologic technologist: Step Go to processing station. This takes approx. The digitizer sends the image data during scanning i.

It checks e. Follow-up period approx. To correct irregularities in the optical path, the digitizer measures at each boot-up the timings of the laser beam between pin diode 1 and 2, 2 and 3 and so on. As a result a kind of "calibration curve" is created, which influences the frequency of AD conversion.

This corrects possible distortions. To correct irregularities in the polygon run, the digitizer measures the timings between the BOL and EOL signal during each scan. Certain irregularities, which may be caused by external vibrations or the polygon ball bearings are smoothed by calculating mean values of several scan lines. When an IP is inserted the polygon starts up. The length of the AD window and "Laser On" depends on the image plate size.

The exact dose is measured by production people and entered in the system. By combining the measured values of several lines the sensitivity is automatically adjusted HT at PMT Shading correction: The digital value of each pixel is corrected with a correction factor determined during shading calibration. How does it work in general: When the stepper motor is accelerating the stall detection is switched off, as torque differs to much during acceleration.

When the motor reaches a stable speed, e. When the torque increases and reaches a certain "switch off torque", as the motor reaches the end position, the motor controller detects the stall position and drives the motor a certain number of steps e. During decelration time the stall detection is switched off, as torque differs to much during deceleration. Rubber pieces in the end position support that the motor can be driven with relatively high speed towards the end position.

If the torque increases to the "switch-off" torque before the motor is supposed to reach the stall position, the motor is also stopped. The software however "knows" that this stop comes too early and issues an error message e.

NOTE: The motor curves shown above are simplified. In reality the motor speed can be reduced just before the stall position, or can even be reduced in several steps before it stops. Supports when a system is planned to be upgraded. The required system components can be analyzed and identified based on the chart. The CR X System version contains a continuous number which is increased when one of the delivered system components changes.

Upgrade possibilities for NX Workstation Software:. NX - Service Bulletin no. CR X Service Bulletin no. Purpose of this Document This document describes the product specific safety notes of the digitizer.

Safety Notes for Installation Planning Safety Notes for Installation Safety Notes for preventive and corrective Maintenance Work General Safety Note for all Service Actvities Safety Note for Replacement of the Power Supply Electrical Check after Power Supply Replacements Intended Use This device must only be used to scan exposed X-ray cassettes, containing an erasable image plate IP.

The digitizer is part of a system, consisting of X-ray cassettes with erasable phosphor image plates and a workstation where the X-ray cassettes are identified and the resulting digital image information is further processed and routed.

It is intended that this device is only operated in a radiological environment by qualified staff. Do not activate cassette edge sensor by hand. NOTE: One sticker on the front cover indicates the correct cassette positioning when inserting the cassette. Remove power plug before servicing. Transport may only be undertaken as described in the operator manual. Vor ffnen des Gertes Netzstecker ziehen. Vor dem Transport unbedingt Hinweise im Bedienhandbuch beachten.

Retirez la prise de courant avant l'entretien. Cat appareil tre transport confrom au manual d'operations. This device complies with Part 15 of the FCC rules.

Operation is subject to the following two conditions: 1 This device may not cuase harmful interference 2 This device must accept any interference received, including interference that may cause undesired operation.

Cet appareil numerique de la classe A respecte toutes les exigences du Reglement sur le material brouillant du Canada. The manufacturer certifies that this product conforms to all applicable provisions of 21 CFR Subchapter J in effect of the date of manufacture. Thi s device complies with Part 15 of the FCC rules. NOTE: The type label is located in the rear.

WARNING: Risk of electrical shock if supplementary earth connector is disconnected with one hand and the other hand touches a ground connection e. First switch off the digitizer and remove the mains cable before you remove the supplementary earth connector. NOTE: With covers closed the digitizer is a Class 1 laser product and not dangerous to the human eye or skin.

Figure 6. Insert the FireWire connector in the rear as indicated on the sticker. Avoid contact with the Erasure Unit. Observe the relevant sticker on the Erasure Unit. Make sure, that the grounding connection to the digitizer frame is present. Figure 9. Avoid direct and indirect eye contact. The structure and stability of the table used, need to be suitable in relation with the size and weight of the system. Electrical shock possible. Ensure that there is either a mains plug or an all-cable disconnecting device in the internal installation fitted near the digitizer and that it is easily accessible.

During transport in a truck for mobile usage and at installations in earthquake areas the digitizer has to be secured by the optional Wall Fixation Kit. See chapter 11, Installation Planning, section "Accessories". Safety Notes for preventive and corrective Maintenance Work This section lists all safety notes for corrective and preventive maintenance.

Risk of live. Mount the cables at the power supply as indicated in the replacement instructions. Handle the light collector with great care. In case of power supply replacements or maintenance check the protective earth resistance according to the national regulations. The electrical resistance has to be 0. Refer to Figure The electrical resistance in this case has to be 0.

Purpose of this Document This document describes all tools software, hardware , which are required for servicing of the digitizer. Reworked procedures for software update at the control PC see section 2. See section 2. Added important note at backup, as it is required at all service interventions to make a backup, except only data are retrieved e. Required Tools Recommended Tools Service Client Menus Service Client Menu Info Counter Service Client Menu E-Label Service Client Menu Error Statistics Service Client Menu Diagnose Cycle Service Client Menu Selftest of Device Service Client Menu Calibration Service Client Menu Stall Calibration Service Client Menu Site specific Data Service Client Menu Device specific Data Service Client Menu Exchanged Components Service Client Menu Maintenance Indicator Machine specific Tools The listing of machine specific tools is divided in required tools and recommended tools.

Required tools: They are required for installation, repair or troubleshooting. Recommended tools: They may facilitate installation, repair or troubleshooting. Required Tools In addition to the conventional service tools which every service engineer has in his toolbox, the following tools are required for service on the digitizer:. Commercially available To remove dust from modules after disassembling from device.

To evaluate the infocounter offline. Commercially available To transfer data between the processing station and the service PC.

NOTE: As some screws are difficult accessible, it is recommended to have a 3 mm hex key with ball-shaped head and a 6 mm magnetic hex key in the tools assortment.

The "Service Client" is the service software for the digitizer. It is installed on the processing station in production. It also contains the software i. Net Visual J to be obtained via www. Depending on the software version, the procedure might be different to the steps described here.

The software installation for the digitizer is made in 2 steps: Step 1: Install the digitizer software incl. Details see this section. Step 2: Upload the software from control PC to the digitizer.

Details see section 2. Otherwise the digitizer driver on the control PC will automatically be de-installed. If at the end of the deinstallation process of the old software a window pops up, telling: "You must restart your system Otherwise continue with next step.

Figure 4 11 When finished, restart the control PC. After de-installation of the old digitizer software on the control PC following error message pops-up:. Restart the control PC. Continue with software installation.

See step 9 previous page. Digitizer was switched not switched off during software de-installation. In this case the digitizer driver on the control PC will also be deinstalled.

Cancel the "Found new Hardware Wizard" and continue with software installation. Install the driver via "found new hardware wizard" which pops up automatically after software upload to the digitizer and restart of the control PC. See Figure 7. Service Client Menus Following menus are available in the digitizer service client:. It facilitates recognition of failures and allows reducing the repair time in the field or the repair center.

Error History with CBF by clicking once on the desired entry in the left pane. The entries of the list are displayed on the right pane. The attributes of the selected entry are displayed in the bottom pane. Service Client Menu E-Label The 'E-labels' 5 pieces altogether; see Figure 9 keep device and module specific data, like module serial number, last error codes or module specific parameters.

Purpose of the E-Labels: o Automatic tracking of module exchange automatic infocounter entries o The returned, defective module has all required data in the E-Label to make fault investigations in production The E-label contains module specific data, avoiding adjustments and loading of hardware modification data at replacements.

Service Client Menu SW-Versioning The software versioning menu allows to check if the software versions currently installed on the processing station match those installed on the digitizer. Service Client Menu Service Activity Log The service activity log menu shows a logfile of all actions carried out by the service staff.

The service person name associated with each logfile entry depends on the CRX service client login. In the example of Figure 11 the login was crservice. Preferably use your name for login. The previous one is automatically renamed to ServiceActivity1. All ServiceActivity. The error statistics menu shows the occurrence frequency for all errors.

It is used for troubleshooting. The menu Hot Report gives a summary of the last cassette cycle. The most important events that happened in the digitizer during this cycle are reported. It is used as a proof during installation, that the digitizer performed a successful IP cycle. The HOT report is overwritten with each new scan cycle. The situation software on digitizer is higher than on control PC may only happen during installation due to logistical reasons.

In this case it is mandatory to install the latest digitizer service client on the control PC. Refer to section 2. Wait for approx. Answer the dialogue popping up did you perform any repair or maintenance activities with "Yes".

Service Client Menu Diagnose Cycle Menu Diagnose Cycle allows troubleshooting image plate handling problems by running partial actions step by step, forward or backward.

This can be helpful when troubleshooting sporadic image plate handling failures. The functional sequence of the selftest is identical to the selftest during switch-on. Refer to chapter 2, Functional Description.

Service Client Menu Calibration The calibration menu allows performing a shading calibration. For details see chapter 3. Service Client Menu Stall Calibration The Stall Calibration menu allows performing a stall calibration for a single motor or all motors. For details see also chapter 3.

NOTE: For troubleshooting of a module the final message parameter iRunCal can be used see Figure 20 : If executed three consecutive times, the displayed iRunCal value may vary 2 digits only e.

Larger deviations show mechanical problems in the mechanics of the respective drive. Menu site specific data allows entering the site specific data for exact identification of the digitizer. The data are part of the infocounter and the E-labels. Menu Device specific data allows to read-out the device specific data which were configured in production.

Menu Exchanged components allows to define whether the exchanged component is a module e. For background information see NOTE next page. Spare part is a submodule of a larger adjusted module and thus not adjusted in production:. Step 2: The service engineer enters in the service client, whether the exchanged part is a submodule of a larger module see scenario 1 above or an own, adjusted module see scenario 2 above. E-Labels contain spare part specific data, entered in production.

To configure whether the screen "Preventive Maintenance required" shall pop-up at the processing station if To make a Backup is required in following cases: At preventive maintenance. After replacements of parts. After software update or upgrade. To make a Restore is required in following cases: After replacement of the processing station.

After replacement of the hard disk on the processing station. After a clean install procedure on the processing station. If these data are lost e. Some data like activity log; site specific data and several infocounter data are not stored on the digitizer and are lost without an up-to-date backup.

These data are however not relevant for proper operation. Without restore the digitizer will boot-up with an error message however, indicating that the serial number of the digitizer is not known. To have all data as up-to-date as possible after exchange of the Control PC or Clean Install Procedure follow the basic rules:. Always make a backup after each service intervention unless only data have been retrieved e.

Remote connection to the processing station and so to the digitizer is performed via Remote Desktop connection. NOTE: The processing station is by default enabled for remote desktop. Purpose of this Document This document describes the meaning of the LEDs and switches on the different printed circuit boards and lists the fuses. ON during reboot only approx. FWI cable connected between digitizer and processing station. Always replace the board if a fuse is defective.

Purpose of this Document This document describes the replacements of parts incl. Where applicable references are made to chapter 3. Removed following replacement instructions and made available as an enclosure to the spare part: o Replacing the Cassette Fixation Steel Band.

Replacing the Front Cover Replacing the Left Cover Replacing the right cover Replacing the Top Cover Replacing the Cassette Edge Sensor Replacing the Cassette Fixation Steel Band Replacing the Cassette Fixation Unit Replacing the Cassette Opener Unit Replacing the Cassette Unlocking Unit Replacing the Chip Reader Replacing the Drawer Unit Replacing the Erasure Unit Replacing the Transport Unit Ball Bearings Replacing the Input Rollers Replacing the IP Handling Board Replacing the Permanent Magnets Replacing the Drawer Motor M Replacing the Fixer Stepper Motor M Replacing the Opener Stepper Motor M Replacing the Roller Stepper Motor M Replacing the Roller IP Sensor Replacing the User Interface Board Replacing the Calibration Board Replacing the Optic Module Replacing the PMI Board Replacing the Scan Rollers Replacing the Slow Scan Board Replacing the Slow Scan Unit Replacing the ESD Brush Replacing the Cassette Unit open Safety Switch Replacing the Distributor Board Replacing the Power Supply Wait till the green light at the user interface lights up takes approx.

Figure 26 13 Remove screw at erasure unit. Do not bend IP guide plate at removal. Figure 32 21 Cut one cable tie at upper part transport unit cable harness left hand side.

Figure 37 26 With a screwdriver remove two ball bearings at transport unit left hand side. Only loosen the screws of the cassette edge sensor approx. Perform a "cassette edge sensor" adjustment. Refer to chapter 3. Figure 57 13 Remove screw at erasure unit.

Figure 63 21 Cut one cable tie at upper part transport unit cable harness left hand side. Figure 67 25 Remove all cables at IP handling board. Figure 68 29 Turn a few turns at spindle of cassette fixation A.

Perform a stall calibration for the cassette opener motor. Cable between chip reader and IP handling board is connected at P Perform a stall calibration for the locking motor. Figure 91 16 Disconnect cable at chip reader. Wait till the status indicator at the digitizer switches to green takes approx. With a screwdriver remove required bearing at right hand side see Figure or left hand side see Figure See NOTE below.

Figure 99 NOTE: As the ball bearings are secured with Loctite, some force has to be applied to remove the ball bearings. For mounting of the ball bearings it is not required to secure them via Loctite again. NOTE: To support that the ball bearing fits in its rail, lift the inner side plate slightly up. It is recommended to replace all 3 magnets together. Figure 12 Remove screw at erasure unit.

Figure 14 Open 3 screws at the fixer stepper motor only 2 screws are shown in Figure Perform a stall calibration for the fixer stepper motor. Figure 12 Remove cable at plug P19 of IP handling board. Perform a stall calibration for the opener stepper motor. Figure 16 Remove cable at plug P25 of IP handling board. Mount in reverse order.

Figure 13 Disconnect two cables at erasure unit. Figure 17 Remove cable at plug P29 of IP handling board. Figure 19 Remove 4 screws of stepper motor roller clamping only 3 screws are shown in Figure Perform a stall calibration for the roller clamping stepper motor. It is not required to replace all toothed belts together. Only replace the defective toothed belt.

Figure 21 Cut one cable tie at upper part transport unit cable harness left hand side. Figure 26 With a screwdriver remove two ball bearings at transport unit left hand side. Perform a stall calibration for the roller clamping motor. Perform a stall calibration for the drawer motor. Perform a stall calibration for the opener motor. When ordering, a different revision of the spare part may be delivered e. It is recommended to replace both scan rollers together.

Figure 13 Remove safety bracket at front scan roller. Perform the cassette unit safety switch position adjustment. Refer to section 3. DANGER: If the power cord is connected and the main switch is switched on the input side of the power supply is connected to high voltage.

Remove all cables in the rear of the digitizer before removal of the rear cover. Figure DANGER: If the cables at the input side of the power supply are mounted wrong, the housing of the digitizer can be connected to high voltage. Mount the cables at the power supply input as indicated on the cables and shown in Figure See section 4. Electrical Check according to National Regulations. Verification of the successful power supply replacement.

Put one probe at the ground connector of the mains input. Put the other probe at a few metallic parts, the customer could touch. Insert the power cord at the digitizer side; leave it disconnected at the wall side.

Put one probe at the ground connector of the mains connector. Purpose of this Document This document describes all adjustments and calibrations as well as their purpose. Added drawer adjustment tool "version 2", which allows to check min. See sections 2. Revised Stall Calibration Procedure: Removed overview table about which stall calibration has to be performed at which replacement, as this is described at each replacement procedure.

Drawer Unit Adjustment Preparing the Drawer Unit Adjustment Checking and Adjusting the lateral Position Checking and Adjusting the Height Checking the Switch Position Adjusting the Switch Position Cassette Edge Sensor Position Adjustment Laser Beam Position Check and Adjustment Performing the Laser Beam Position Check Performing the Laser Beam Position Adjustment Shading Calibration Verifying a successful Shading Calibration Possible Reasons for a failed Shading Calibration Stall Calibration Overview of Adjustments and Calibrations The following tables give an overview of adjustments and calibrations for the digitizer.

In case of sporadic or persistent errors or to identify the possible root cause of an incorrect laser beam position. To solve errors or , if an incorrect laser beam position is the root cause. Whenever the PMT or light collector have been removed e. Whenever stripes in slow scan position are visible, which cannot be removed by cleaning of the scan line with the cleaning brush.

After replacement of a motor or mechanical assembly where a motor with stall detection is built in After replacement of the IP handling board as this has the motor controllers built in During preventive maintenance.

Adjustment of the optimum height and lateral position of the upper drawer unit plate for reliable image plate transport in and out of the cassette. Difference: Drawer adjustment tool version 2 has an additional gauge integrated to check the maximum height position marked with text "max.

The drawer adjustment tool is part of delivery of following spare parts:. This document describes the adjustment with drawer adjustment tool version 2. Where applicable, differences to version 1 are described. Simultaneously press the two buttons A positioned underneath the feed table. When pushing the drawer unit only do this via drawer unit axle. Check, whether plastic finger at drawer unit is aligned with the lateral stop of the drawer adjustment tool.

In case the lateral stop is aligned with the plastic finger at the drawer unit there is no adjustment required. Continue with section 2. Otherwise perform lateral adjustment as described in next steps. The plate marked with "min. The plate marked with "max. NOTE: With drawer adjustment tool version 1 it is possible to check the minimum height only. It is not possible to check the maximum height.

In case the height is OK on both sides there is no height adjustment required. Otherwise proceed with next steps. Press down the drawer unit upper plate after each adjustment cycle before checking the height again. Again, check the height adjustment min. Wait till the green light at the user interface lights up takes approximately 1 minute. Miss-alignment can lead to IP jams, laser reflections or cracks in the light collector. If the distance is OK, continue with step 9 , otherwise continue with next step.

See Figure The PMI board carrier is adjusted this way, that the light collector is mounted stress free. Check: With closed cassette unit the switch may not interrupt the power: The status indicator LED lights up.

Figure Check: With cassette unit opened approximately 1 cm, the switch must switch off the digitizer. If it does not switch off, continue with section 2. Shift the switch to the front switch does not close with inserted cassette unit or to the rear switch switches too late when cassette unit is opened.

Cassette Edge Sensor Position Adjustment Adjustment of the switch position, so that it reliably detects whether the cassette is inserted the right way. After replacement of the cassette edge sensor. Insert a cassette the right way, i. If the sensor does not switch, open the fixation screws at the cassette edge sensor and push it more to the left. Fasten the fixation screws and repeat step 2. Insert a cassette the wrong way, i.

When using a multimeter, the resistance should be outside the measuring range. If the sensor switches, open the fixation screws at the cassette edge sensor and push it more to the right. Repeat step 5. If the sensor does not switch, repeat the complete procedure once more, to confirm the proper adjustment. Adjustment of the optic module this way that the laser beam position is centered to the calibration board.

In case of sporadic or persistent errors or to identify the possible root cause of an incorrect laser beam position To solve errors or , if an incorrect laser beam position is the root cause. By putting the switch on the distributor board to ON position the laser beam position check functionality is activated: The yellow "LED 1" lights up.

Do not touch the input side of the power supply by hand or with any tool. Check that the yellow LED on the distribution board is ON: This is the proof, that the check functionality is activated. Eye injuries possible. Do not keep any tool in the laser beam while the laser is switched on. The Polygon motor is audible.

NOTE: The laser beam switches off automatically after 30 minutes. Check, whether laser beam is really switched on: Look from the digitizer right hand side, whether the red laser line is visible. If laser beam is visible, the laser beam position adjustment is required.

Set the switch on the distributor board to the OFF position before handing the digitizer over to the customer. The Laser Beam Position Check is performed.

See steps 1 to 17 in section 2. Open the four screws at the optic module position plates secured with red lacquer. If the laser beam position adjustment has been performed: Verification is done with the shading calibration procedure. Determine the sensitivity of the scanning system for each pixel of the scan line: A flatfield is scanned. Perform a shading calibration in following cases:. Select the largest available cassette. Open the cassette and check the image plate IP for scratches.

If no scratches are visible, select this IP. Ensure that the entire image plate is fully exposed! The collimated field must be larger than the image plate. After the second exposure the cassette has been exposed in total with a dose of approximately 20 Gy.

In case of an error refer to section 3. To keep a certain calibration file, save it with a different name. If the scanned image does not show stripes in fast scan or slow scan direction or large area inhomogeneities no further action is required. In such a case please contact your local Agfa representative. Checking the images may only happen on light boxes and viewing stations that comply with the specified installation and configuration conditions for diagnostic reading.

For details please refer to the respective instructions of the viewing stations or light boxes. Possible Reasons for a failed Shading Calibration There are mainly following reasons for a failed shading calibration:. Reason for different proceeding: In opposite to Genrad General Radiology cassettes, FLFS cassettes do not have the complete IP area covered with a backscatter protection: The lead foil does not cover approximately 10 mm on the upper and lower side of the cassette.

NOTE: This covers the area where the cassette has no backscatter protection. Only this area has to be covered, which is used by the digitizer to determine the shading calibration curves. FLFS cassette. Put a cassette underneath this area. This prevents, that the missing backscatter protection falsifies the shading calibration result.

These darker zones also have to be used for image quality evaluation in slow scan and fast scan direction to verify a successful shading calibration. The stall calibration determines the optimum motor current for stall detection: The motor drives several times to its home and end position with different motor currents. The mean value of these different motor currents is used as motor control current, where the steps of the motor are in the specified, allowed range.

It compensates the differences e. After replacement of a motor or mechanical assembly where a motor with stall detection is built in. NOTE: The stall calibration does not harm - it can be performed anytime.

It is required however only in the cases as described previous page. The digitizer CRX does not have device specific settings which may be altered by the service engineer.

All settings are either determined in production and stored in E-Lables in the corresponding module, or - at replacement of parts - have to be determined again by an adjustment routine. These new values are automatically written to the corresponding E-Label. Purpose of this document This document lists all customer released software versions with their features, limitations and known problems. Supported Functions New features for the Customer New features for Service Additional, minor Bug Fixes Samsung Combination Firmware is special service firmware.

You can flash the device with this Combination Rom in order to check all hardware features like back camera, front camera, sensors, RAM memory, ROM memory. You may also run a test that shows you the results of the phone diagnostics. In the table below we present the list of all combination firmware that you can download.

WebAssembly is a technology that is going to revolutionize the web as we know it. Speed up your web applications, or put your software on the web with ease! In this article, we'll show popular IP addresses. These are some of the most used and well-known IP addresses you can use in your computer network! PPPoE is a great protocol that allows us to basically use the internet on a daily basis without any problems.



0コメント

  • 1000 / 1000